Experiment Session Notes
- Mouse ID: 794237
- Date: 2025-04-03
- Experimenter: @Seyedolmohadesin @rcpeene @jeromelecoq @KasparP
- Rig / Setup ID: SLAP2
- Stimulus version: 30a4291398f5054fcb769db5d45e388f34dcbce2
- Protocol followed: Glutamate imaging, 30 min experiment
- Notes & Issues:
- Virus details:
- (pHp.eB) 5E12 ExP32222-hSyn-Flex-iGluSnFR4.v9601.NGR
- 1E12 AAB.pHp.eB-CAG.Flex.NES.JRGECO1a.WPRE.SV40
- 6E8 AAV.php.eB-CamK|I0.4.Cre.SV40
- Bonsai files were changed to have clean START and STOP logging in CSV files.
- CSV also contains START and STOP of SLAP2 as well as SPACE BAR frame number.
- The DIO2 port is turned on when grating starts to provide accurate alignment to grating starts.
- Stimulus has a full receptive field mapping code now pulling from a separate CSV file for all parameters.
- Mouse:
- C57 wild type, male, born 2025-01-07.
- Injected with iGluSnFR4 in V1 on 2025-03-12.
- Imaging:
- Started at 10:39, ended at 11:13.
- Input to BONSAI:
- Baseline: 960
- Mismatch: 8
- Receptive field: 6
- We should check code for Baseline/Mismatch ratio to guarantee the right occurrence probability.
- Two planes chosen: Layers 2/3 cell.
- We are approximately in V1, this is not precisely targeted in this experiment.
- One soma and proximal dendrites in upper layers.
- Planes are chosen by eye.
- Dendrites are drawn by hand in proximal and apical dendrites.
- Signal was strong, somas and dendrites clearly visible.
- Maedeh acquired reference stacks for online motion correction.
- Laser cooling oscillating fan caused noise every 20s (measured with watch), probably related to some feedback weight that needs to be optimized.
- We expect this sound to be noticeable by mice.
- Data stored at:
/allen/scratch/OpenScope/Slap2/Data/794237/
- Frame logging CSV file ended 2 min in due to a bug.
- We identified the bug in the code (a TakeUntil was left with a delay).
- The grating should still be possible to align since we have all parameters of the orientations.
- A pulse is sent on Digital Port 2 of the Bonsai card at the start of each grating along with start and stop of SLAP2 imaging on DO0 and D01.
- We confirmed in the file that those are present.